5 dc power source 25, 6 power on self-test 25, 7 display backlight and contrast 25 – Fluke 1529 Chub-E4 User Manual

Page 32: 8 taking measurements 25, 1 connecting the sensor 25, Backlight 25, 35, 53, Dc power 25, Self-test 25, 140, Dc power source, Power on self-test

Advertising
background image

6.5

DC Power Source

Caution: For CE compliance and for performance, use only the AC

adapter shipped with the instrument by Hart Scientific. If the AC adapter
needs to be replaced, contact a Hart Authorized Service Center (see Sec-
tion 1.3).

The DC power source provides power to charge the battery. It can also be used
to power the 1529 while the battery is being charged. The AC adapter provided
with the 1529 is intended for these purposes. The DC power source plugs into
the DC power input on the back panel of the instrument. The AC adapter has
circuits with high voltages inside that could present danger of electric shock or
fire if exposed. If the AC adapter is damaged in any way or becomes hot, dis-
continue use immediately, disconnect the adapter from any AC supply, and re-
place the adapter. Do not attempt to open, repair, or continue using a damaged
or defective AC adapter.

6.6

Power On Self-Test

When power is turned on, the 1529 performs a self-test checking the system, all
channels, calibration, GPIB, memory, and buttons. If an error occurs, an error
message is displayed. See Section 11.1, Troubleshooting for additional infor-
mation on error messages.

6.7

Display Backlight and Contrast

The display backlight is adjustable for use in varying lighting conditions. The
four backlight modes are accessible in the CHANNEL menu DISPLAY OP-
TIONS function. If the display appears faded, dark, or blank, adjust the con-
trast (off, low, medium, high) using the front panel contrast buttons.

6.8

Taking Measurements

The procedure for configuring the 1529 thermometer readout to take measure-
ments on a particular input channel requires the following: (1) connecting the
sensor to the appropriate input channel , (2) enabling the channel, and (3) se-
lecting the conversion type and probe characterization values.

6.8.1

Connecting the Sensor

Connect the sensor to the RTD/Thermistor channel(s) or thermocouple chan-
nel(s) that you want to measure. Refer to Section 7.4, Probe Input Modules, for
additional information.

25

6 General Operation

DC Power Source

Advertising