5 format:stamp? 90, 6 format:stamp <bool> 90, 7 measure? [<chn>] 90 – Fluke 1529 Chub-E4 User Manual

Page 96: 8 read? [<chn>] 90, 9 sense<chn>:data? 90, Format:stamp, Format:stamp <bool, Measure? [<chn, Read? [<chn, Sense<chn>:data

Advertising
background image

day the measurement was made

With this format kohms and millivolts are not converted to ohms and volts.

This command may return the same reading if a new reading has not yet been
obtained. If only new readings are desired, the FORMat:STAMp command can
be used to enable the extended format option and any readings in which the
value of the first field is 0 can be ignored or else the STATus:MEASure? com-
mand can be used to determine if a new reading is available for a desired
channel.

8.4.1.5

FORMat:STAMp?

Returns the extended measurement data format setting. See Section 9.4.1.4,
FETCh? for details.

FORM:STAM?

0

8.4.1.6

FORMat:STAMp <bool>

Enables extended measurement data format with the FETCh?, MEASure? And
READ? commands. On power up this is always set to off.

FORM:STAM ON

8.4.1.7

MEASure? [<chn>]

Returns the most recent measurement for the specified channel.

MEAS? 1

0.0127

This command is the same as the FETCh? [<chn>] command

8.4.1.8

READ? [<chn>]

Returns the most recent measurement for the specified channel.

READ? 2

0.0113

This command is the same as the ?FETCh [<chn>] command.

8.4.1.9

SENSe<chn>:DATA?

Returns the input value for the specified channel.

SENS3:DATA?

100.0291, 0.0000

1529 Chub-E4 Thermometer Readout
User’s Guide

90

Advertising