8.4.7.8

LOGging:AUTomatic:PRINt [<num>|ALL [,<port>]]

Prints the auto log entries.

LOG:AUT:PRIN ALL

DATA_25 1 22.676C 12:19:42 09-05-00
DATA_25 2 9.960 KO 12:19:44 09-05-00
DATA_25 3 23.220C 12:19:46 09-05-00
DATA_25 4 23.245C 12:19:48 09-05-00
DATA_25 1 22.765C 12:19:52 09-05-00
DATA_25 2 9.962 KO 12:19:54 09-05-00
DATA_25 3 23.087C 12:19:56 09-05-00
DATA_25 4 23.260C 12:19:58 09-05-00...

If no parameters are specified, the ALL parameter is assumed and all auto log
entries are printed to the serial port. Specifying the <num> parameter prints
only the log entries with the specified label number to the serial port. Op-
tionally, the log entries may be printed to a specific port by specifying the
<port> parameter (see Table 10 on page 129, Port Numbers).

8.4.7.9

LOGging:AUTomatic:STATus?

Returns the auto log status.

LOG:AUT:STAT?

0

If this query returns a 0, an auto log session is not running. If this query returns
a 1, an auto log session is running.

8.4.7.10

LOGging:AUTomatic:STATus <bool>

Sets the auto log status.

LOG:AUT:STAT 0

The <bool> parameter turns auto logging on (1 or ON) or off (0 or OFF). If si-
multaneous scan mode is enabled, one measurement is stored for each channel
in the auto log. The *RST command set the auto log status to off.

8.4.7.11

LOGging:AUTomatic:TIMe? [MIN|MAX|DEF]

Returns the auto log interval setting.

LOG:AUT:TIM?

1

Specifying the MIN parameter returns a value of 0.1. Specifying the MAX pa-
rameter value returns value of 3600. Specifying the DEF parameter returns a
value of 1. The units are seconds.

1529 Chub-E4 Thermometer Readout
User’s Guide

108