3 password 74, Figure 45 password 74, Password 74 – Fluke 1529 Chub-E4 User Manual

Page 80: Password, Figure 45, 159c, 142c, 634c, 015c, 3 password

Advertising
background image

The D FORM parameter allows the user to set the date format used when dis-
play the date. The choices are M-D-YY, M-D-YYYY, D/M/YY, and
D/M/YYYY.

Press the

LR buttons to change the parameter setting. Use the UD buttons

to scroll between the parameters. Press the Enter button to save the new set-
ting. Press the Exit button to exit to the menu.

7.5.3

Password

The PASSWORD function is used for changing the password required for ac-
cess to password-protected parameters and to determine whether to pass-
word-protect certain parameters. When this function is selected, the bottom
portion of the display prompts the user to enter the current password to gain ac-
cess to this function.

By default the password is set to '1529' when the instrument ships from the fac-
tory. We recommend that the user change the default password to protect the
password-protected settings. Note: Keep your password in a secure location
and do not forget it.

To enter the password, press the Enter button. Press the

LR buttons to high-

light the digit you want to change. Use the

UD buttons to scroll through the

numbers 0-9 to set each digit of the password. Press the Enter button to accept
the password or press the Exit button to cancel and to exit to the menu.

If an incorrect password is entered, the display shows the message PASS-
WORD INCORRECT and access to the password-protected parameters is de-
nied. Press the Enter button to return to the menu. If the correct password is
entered, the password-protected parameters are displayed in the bottom portion

1529 Chub-E4 Thermometer Readout
User’s Guide

74

1

24.159C

3

24.142C

2

24.634C

4

24.015C

ENTER CORRECT PASSWORD FOR ACCESS.

PASSWRD:

0

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 45 Password

Advertising