2 enable channel 32, Figure 9 enable channel 32, Selection 32 – Fluke 1529 Chub-E4 User Manual

Page 38: Enable channel 32, Enable channel, Figure 9, 159c, 142c, 634c, 015c

Advertising
background image

Press ENTER to continue …

Press the Enter button to confirm and to continue. The display is remains un-
changed and all enabled channels remain enabled channels. Press the Exit but-
ton to cancel.

When the measure period is reset to 1 second or greater, the following message
is displayed indicating that the 1529 is returning to normal operation.

FULL ACCURACY.

The DISPLAY FUNCTIONS function of the CHANNEL menu reappears and
the CHANNEL MODE function is available.

Note: Accuracy in the Fast Measurement Mode is affected by ambient condi-
tions, changes in measured temperature, higher measured temperature, and the
time duration from using the AUTO-CAL function. To reduce the uncertainty
in the Fast Measurement Mode, use the 1529 in a controlled ambient environ-
ment, in a very stable bath or dry block at lower temperatures, and use the
AUTO-CAL function often.

7.1.2

Enable Channel

The ENABLE CHANNEL function enables and disables measuring of each
channel. If a channel is off, its measurement is not displayed on the top portion
of the display in either scan or simultaneous display modes. When this function
is selected, the bottom portion of the display shows the on/off status of each
configured channel and allows the setting to be changed. Each configured chan-
nel is identified by its channel number.

Use the

UD buttons to select the channel to change. The LR buttons are

used to change the setting. Press the Enter button to save the setting. Press and

1529 Chub-E4 Thermometer Readout
User’s Guide

32

1

24.159C

3

24.142C

2

24.634C

4

24.015C

CHAN 1:

ON

CHAN 2:

ON

CHAN 3:

ON

CHAN 4:

ON

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 9 Enable Channel

Advertising