1 measurement commands 88, 1 calculate:average:clear 88, 2 calculate<chn>:average<n>:data? 88 – Fluke 1529 Chub-E4 User Manual

Page 94: Measurement commands, Calculate:average:clear, Calculate<chn>:average<n>:data, 1 measurement commands

Advertising
background image

• <num> indicates numeric value is required.
• <chn list> indicates a list of channel numbers is required.
• <bool> indicates a Boolean value (0 or 1) is required. The mneumonics

OFF and ON are also accepted for 0 and 1, respectively.

• <conv> indicates a conversion mnemoic is required.
• <param> indicates a parameter name is required.
• <seri> indicates a serial number is required.
• <res> indicates a resistance value is required.
• <volt> indicates a voltage value is required.
• <unit> indicates a temperature unit is required.
• <pass> indicates a password is required.
• <port> indicates a port number is required.
• <label> indicates an eight character label is required.
• <year> indicates a four digit number is required.
• <month> indicates a one or two digit number is required.
• <day> indicates a one or two digit number is required.
• <hour> indicates a one or two digit number is required.
• <minute> indicates a one or two digit number is required.
• <second> indicates a one or two digit number is required.
• <baud> indicates a valid baud number is required.

8.4.1

Measurement Commands

The measurement commands are used to read information on the statistical cal-
culations and for reading the most recent measurement for a specified channel.
The measurement commands are not password protected.

8.4.1.1

CALCulate:AVERage:CLEar

Clears the statistical functions for all channels.

CALC:AVER:CLE

The *RST command clears the statistical functions.

8.4.1.2

CALCulate<chn>:AVERage<n>:DATA?

Returns the value of a statistical calculation for a given channel.

CALC3:AVER1:DATA?

0.017

1529 Chub-E4 Thermometer Readout
User’s Guide

88

Advertising