2 *opt? 112, 3 *rst 112, 4 system:boot:version? 112 – Fluke 1529 Chub-E4 User Manual

Page 118: 5 system:code:version? 112, System:boot:version, System:code:version

Advertising
background image

*IDN?

HART,1529,A09001,1.11

8.4.8.2

*OPT?

Returns a list of input modules installed.

*OPT?

PRT,TC

Input module types are reported by position, top to bottom, of the two input
modules. The returned example is for a 1529 equipped with one 2-channel PRT
input module and one 2-channel Thermocouple input module.

8.4.8.3

*RST

Sets the instrument operating parameters to defined conditions.

*RST

The following parameters are set when this command is issued.

• Sets the scanning sequence timer to 1 second.
• Sets channel 1 ON and all others OFF.
• Sets the scan mode to SCAN.
• Sets the moving average to 1.
• Sets the printing to all ports OFF.
• Sets the units to C.
• Sets the resolution to AUTO.
• Stops auto logging.
• Clears the statistical functions.

8.4.8.4

SYSTem:BOOT:VERSion?

Returns the boot ROM version.

SYST:BOOT:VERS?

0.50

8.4.8.5

SYSTem:CODE:VERSion?

Returns the firmware code version.

SYST:CODE:VERS?

1.11

1529 Chub-E4 Thermometer Readout
User’s Guide

112