proving resolution. When this function is selected, the bottom portion of the
display shows the averaging setting and allows the setting to be changed. Select
the number of raw measurements to be averaged to produce the displayed mea-
surement. The range is from 1 to 10. The factory default is 1.

The measurements are averaged together until the selected number of measure-
ments to average has been reached. Then the calculated average is based on the
selected number of measurements to average. For example, if the selected num-
ber of measurements to average is selected as 3, the 1st measurement is dis-
played, the 1st and 2nd measurements are averaged and displayed, and then the
1st, 2nd, and 3rd measurements are averaged and displayed. From this point
forward each displayed measurement consists of the last three measurements
averaged together.

The

LR buttons are used to select a setting. Press the Enter button to save

the setting and exit. Press the Exit button to cancel and to exit to the menu.

7.1.5

Display Options/Auto-Cal

The DISPLAY OPTIONS function is displayed when the measure period is one
second or greater and allows you to select the display resolution, decimal char-
acter, and lamp illumination. The AUTO-CAL function is displayed when the
measure period is 0.1, 0.2, or 0.5 seconds and allows a full measurement to be
taken. When the DISPLAY OPTIONS function is selected, the bottom portion

1529 Chub-E4 Thermometer Readout
User’s Guide

34

1

24.159C

3

24.142C

2

24.634C

4

24.015C

AVERAGE:

3

EXIT

ENTER / MENU

UNIT

CONTRAST

1529

CHUB E-4 THERMOMETER READOUT

Figure 11 Moving Average