13 status:operation? 124, 14 status:operation:condition? 124, 15 status:operation:enable?[min|max|def] 124 – Fluke 1529 Chub-E4 User Manual

Page 130: 16 status:operation:enable <num>|min|max|def 124, 17 status:questionable? 124

Advertising
background image

STAT:MEAS:ENAB 15

8.4.12.13

STATus:OPERation?

Reads and clears the Operation Status Register.

STAT:OPER?

16

The assignment of the bits in this register is as follows:

bits 0-3: not used

bit 4: measurement complete

bits 5-15: not used

8.4.12.14

STATus:OPERation:CONDition?

Returns the Operation Status Condition Register.

STAT:OPER:COND?

16

The assignment of the bits in this register is as follows:

bits 0-3: not used

bit 4: measurement in process

bits 5-15: not used

8.4.12.15

STATus:OPERation:ENABle?[MIN|MAX|DEF]

Returns the Operation Status Enable Register.

STAT:OPER:ENAB?

255

8.4.12.16

STATus:OPERation:ENABle <num>|MIN|MAX|DEF

Sets the Operation Status Enable Register.

STAT:OPER:ENAB 16

8.4.12.17

STATus:QUEStionable?

Reads and clears the Questionable Status Register.

STAT:QUES?

16

The assignment of the bits in this register is as follows:

bits 0-3: not used

1529 Chub-E4 Thermometer Readout