9 communication interface commands 114, 1 system:communicate:ir:feed? [min|max|def] 114, 3 system:communicate:ir:mode? [min|max|def] 114 – Fluke 1529 Chub-E4 User Manual

Page 120: 4 system:communicate:ir:mode <num>|min|max|def 114, Communication interface commands, System:communicate:ir:feed? [min|max|def, System:communicate:ir:feed <bool>|min|max|def, System:communicate:ir:mode? [min|max|def, System:communicate:ir:mode <num>|min|max|def, 9 communication interface commands

Advertising
background image

8.4.9

Communication Interface Commands

The communication interface commands are used for querying and setting se-
rial port, IR, and IEEE configuration parameters.

8.4.9.1

SYSTem:COMMunicate:IR:FEED? [MIN|MAX|DEF]

Returns the state of the IR port printing.

Example: SYST:COMM:SER:FEED?

Response: 0

A value of 1 is returned if IR port printing is enabled. A value of 0 is returned
if IR port printing is disabled. Specifying the MIN parameter returns a value of
0. Specifying the MAX parameter returns a value of 1. Specifying the DEF pa-
rameter returns a value of 0.

8.4.9.2

SYSTem:COMMunicate:IR:FEED <bool>|MIN|MAX|DEF

Sets the state of the IR port printing.

Example: SYST:COMM:SER:FEED DEF

The <bool> parameter turns IR printing on (1 or ON) or off (0 or OFF). Spec-
ifying the MIN, MAX, or DEF parameter sets the IR printing to the minimum,
maximum, or default value respectively. The *RST command sets the IR port
printing to OFF.

8.4.9.3

SYSTem:COMMunicate:IR:MODE? [MIN|MAX|DEF]

Returns the IR mode setting.

Example: SYST:COMM:IR:MODE?

Response: 0

A value of 0 is returned if the IR mode is off (IR port is disabled). A value of 1
is returned if the IR port is in HP printer mode. A value of 2 is returned if the
IR port is in communications mode. This mode should be used when using the
thermometer readout with the LogWare software. A value of 3 is returned if the
IR port is in IRDA mode. Specifying the MIN parameter returns a value of 0.
Specifying the MAX parameter returns a value of 2. Specifying the DEF pa-
rameter returns a value of 0.

8.4.9.4

SYSTem:COMMunicate:IR:MODE <num>|MIN|MAX|DEF

Sets the IR mode setting.

Example: SYST:COMM:IR:MODE 1

The <num> parameter specifies the IR mode setting. A value of 0 turns IR
mode off (IR port is disabled). A value of 1 sets the IR mode to HP printer
mode. A value of 2 sets the IR mode to communications mode. A value of 3

1529 Chub-E4 Thermometer Readout