Creating an isotherm temperature band – Fluke Ti50 User Manual

Page 50

Advertising
background image

Ti40, Ti45, Ti50, Ti55

Users Manual

4-8

Creating an Isotherm Temperature Band

Note

This feature is available only on Ti45 and Ti55 cameras without
IR Fusion.

You can add an isotherm temperature band to an image to highlight areas of
equal temperature. The image pixels that fall within the isotherm span you set
appear in a color different from the palette colors. You can specify the span
using either the maximum and minimum temperatures or the center point
temperature and span width.

To create an isotherm temperature band:

1. Tap F.

2. Use the MOUSE controller to position the pointer over Image Settings on

the pop up menu.

3. Tap E.

4. Position the pointer over the Iso. tab and tap E.

5. Position the pointer over the Enable Isotherm box and tap E. A check

mark indicates the feature is enabled.

6. Position the pointer over either the Max/Min or Center/Width left/right

and tap E repeatedly as needed to increase/decrease the isotherm span
settings.

7. Tap the TRIGGER button to accept setting change and return to scan

target mode.

Note

The isotherm’s temperature limits are indicated next to the color
palette on the display screen. To adjust these isotherm temperature
limits:

1. Position the pointer on over the isotherm temperature limit you want

to change, then tap E to highlight the value.

2. Use the MOUSE controller up/down action to change the value.

3. When you reach the desired temperature, tap the TRIGGER button to

accept setting change and return to scan target mode.