Mpj™+, psr+, psf+ joysticks or displays – Invacare TDX SP User Manual

Page 32

Advertising
background image

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

TDX™ SP

32

Part No 1143190

Service Indicator Light Diagnostics

MPJ™+, PSR+, PSF+ Joysticks or Displays

NUMBER

OF

FLASHES

ERROR CODE

DESCRIPTION

POSSIBLE SOLUTION

1

User Fault

Release joystick to neutral and try again.

2

Battery Fault

Charge the batteries. Refer to Charging Batteries on page 62.
Check that battery cables are connected properly. If necessary,
replace batteries. Refer to Replacing Batteries on page 55.

3

Left Motor Fault

Contact Invacare/Dealer for service.

4

Right Motor Fault

Contact Invacare/Dealer for service.

5

Left Park Brake Fault

Contact Invacare/Dealer for service.

6

Right Park Brake Fault

Contact Invacare/Dealer for service.

7

Remote Fault

Check to make sure joystick is connected properly.
Contact Invacare/Dealer for service.

8

Controller Fault

Contact Invacare/Dealer for service.

9

Communications Fault

Contact Invacare/Dealer for service.

10

General Fault

Contact Invacare/Dealer for service.

11

Incompatible or incorrect

Remote

Wrong type of remote connected. Contact Invacare/Dealer
for service.

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

ƽ

SPM L Park Brake Fault or

ƽ

SPM R Park Brake Fault displays

and wheelchair does not drive.

Motor lock levers disengaged
(Error code E9 or E10).

Engage motor lock levers. Refer to
Disengaging/Engaging Motor Lock
Levers on page 48.

CHARGER PLUGGED IN displays.

Battery charger connected
(Error code E28).

Unplug battery charger from the
wheelchair. Refer to Charging Batteries
on page 62
.

ƽ

SPM Battery Fault displays and the

wheelchair does not drive.

Batteries need to be charged
(Error code E14).

Charge batteries. Refer to Charging
Batteries on page 62. If
batteries fail to
charge properly, check battery charger
or replace batteries. Refer to Replacing
Batteries on page 55.

JOYSTICK TIMEOUT displays and

the wheelchair does not drive.

Joystick or input device is
disconnected (Error code 32).

Turn off power, reconnect the joystick
of input device and turn power on.

JS REV TOO LARGE

JS FWD TOO LARGE

JS LFT TOO LARGE or

JS RGT TOO LARGE

displays and the wheelchair does not
drive.

The joystick or input device is
sending a value outside of the
reverse, forward, left or right
limits (Error codes E01, E02,
E03 or E04).

Replace joystick or input device.

Advertising