Section 9—forks, Adjusting forks – Invacare TDX SP User Manual

Page 52

Advertising
background image

SECTION 9—FORKS

TDX™ SP

52

Part No 1143190

SECTION 9—FORKS

ƽ WARNING

After ANY adjustments, repair or service and before use, make sure that all

attaching hardware is tightened securely - otherwise injury or damage may result.

CAUTION

As with any vehicle, the wheels and tires should be checked periodically for cracks

and wear, and should be replaced.

Adjusting Forks

NOTE: For this procedure, refer to FIGURE 9.1.

1. Remove the retaining screw that secures the headtube cover in place.

2. Remove the headtube cover (not shown) from the caster headtube.

3. To properly tighten caster journal system and guard against flutter, perform the 

following check:

A. Tip back the wheelchair to floor.

B. Pivot both forks and casters to top 

of their arc simultaneously.

C. Let casters drop to bottom of arc 

(wheels should swing once to one‐
side, then immediately rest in a 
straight downward position).

D. Adjust locknuts according to 

freedom of caster swing.

FIGURE 9.1 Adjusting Forks

4. Test wheelchair for maneuverability.

5. Readjust locknuts if necessary, and repeat STEPS 1‐3 until correct.

6. Snap headtube cover into the caster headtube.

7. Reinstall retaining screws.

Locknut

Fork

Retaining Screw

Caster Headtube

Headtube Cover