Chapter 6 Unit Maintenance

HRG010-W HRG015-W

6.2 Inspection and Cleaning

1st edition : Aug. 2006
Rev. A: May. 2007

6-3

6.2.1 Daily

inspection

Perform daily inspection of the items listed in the Table 6-2. If the check finds any
abnormal event, stop system operation and turn off the main power supply.
Request for service promptly.

Table 6-2 Daily Inspection

Inspection item

Inspection method

No heavy object is placed on this unit. This unit
should not be subjected to external force.

Installation
condition

Check of the condition of
unit installation

Temperature and humidity fall within the specified
range.

Fluid leak

Check of the piping
connector section

No leak of facility water and circulating fluid from the
piping connector sections

Water supply

Confirm the reading on the
level indicator

The fluid level falls within the specified range.

Display check

The diplsay assures sharp image and numbers.

Operation panel

Function check

The lamp is ON.

Circulating fluid
temperature

Confirm the reading on the
LCD screen

Normal temperature should be assured for system
use.

Circulating fluid
outlet pressure

Confirm the reading on the
circulating fluid pressure
gauge

Normal pressure should be assured for system use.

Refrigerant high
pressure

Confirm the reading on the
refrigerant high pressure
gauge

Normal flow rate should be assured for system use.

Operating
condition

Operating condition check

No abnormal noise, vibration, odor and smoke

Facility water

Check of facility water
condition

Temperature, flow rate, and pressure fall within the
specified range.

If the above check finds any abnormal event, bring the unit to a halt and turn off
the main breaker. Make sure to perform lockout and tagout before requesting
technical service.