Scale Setting (SCL Setting)

To assign a different scale to the measured data, the range can be set as described
below. Note that the measured data which can get a different scale must be of the
voltage (VOLT). To assign a unit to this new scale, see section 6.4.

Key operation

Indication

Discription

Press the [DISP] key for three seconds to enter the SET mode.
Select the [

] display by using the [

] key.

Then press the [ENT] key.

Use the [

] key to select the desired channel.

Press the [ENT] key.

Use the [

] key to select [

].

Press the [ENT] key.

Select the desired range (using the [

] key) and press the

[ENT] key.

The display which appears will allow you to enter the
minimum value of the recording span (SPAN L). Select the
desired minimum value using the [

] and [

] keys. Then

press the [ENT] key.

The display which appears will allow you to enter the
maximum value of the recording span (SPAN R). Select it the
same way as the minimum value and press the [ENT] key.
Note that SPAN L and SPAN R cannot be same.

This display allows you to enter the minimum value of the scale
(SCL L) corresponding to the minimum value of the recording
span (SPAN L). Select it using the [

] and the [

] keys. The

position of the decimal point can be moved using the [

] key

in case [

] is flashing. Note that the allowable range is –19999

to 20000. Then press the [ENT] key.

This display allows you to enter the desired maximum value of
the scale (SCL R) corresponding to the maximum value of the
recording span (SPAN R). Select it the same way as the
minimum value. Then press the [ENT] key. Note that the
decimal point will be in the same position as for SCL L. SCL L
and SCL R cannot be same.

[

] is displayed and the setting is completed.

Press the [ENT] key to set another channel; or press the
[DISP] key for three seconds to return to the OPERATION
mode.

NOTE

• In this example, 0.000V input will be scaled to –100.0. 2.000V input will be

scaled to 100.0.

• If you press the [ESC] key in the middle of setting, the indication returns to the

[

] and already set data will be aborted.