6. SET MODE

6.2

Setting Offsets (OFFST)

When recording and showing data using the current input as the reference, measured
values can be corrected within 10% of the span. For example, assume the reading is
1.998 V for the current level of input. If you want to change the reading to 2.000 V, set
0.002 V as the offset. This menu opens only when offset settings protection is turned
on in the SET UP mode.

Key operation

Indication

Discription

Enter the SET mode by pressing the [DISP] key for three
seconds. Using the [

] key, show [

] and then

press the [ENT] key.

Using the [

] key, select the channel to be configured and

then press the [ENT] key. Configurable channels are those
whose measuring range is VOLT, TC, RTD or SCL.

Type the value of offset. Use the [

] key to move through

the digits and the [

] key to select a numeral. When setting

is complete, press the [ENT] key. Note that set the offset
within 10% of the recording span or scale.

[

] is displayed and the setting is then completed.

Press the [ENT] key to set another channel; or press the
[DISP] key for three seconds to return to the OPERATION
mode.

NOTE

If you press the [ESC] key in the middle of setting, the indication returns to the

[

] and already set data will be aborted.