Omega RD260A User Manual

Page 94

Advertising
background image

8 - 4

Does input signal satisfy

the specifications?

NO

NO

NO

NO

YES

YES

YES

¥ Errors exceed specified limits

¥ Measuring value indications fluctuate

¥ Pen/printer carriage goes off-scale at 0% or 100%.

YES

Correct the input according to specifications

Are input wires correctly

connected to the terminals?

¥ Correctly attach the terminal cover.
¥ Check the wind from appliances such
as cooling fan.
¥ Decrease the temperature change
around the input terminals.

Is the input free of noise?

Are measures applied

to prevent ambient

temperature changes?

¥ Separate the input signal wires from
the noise source.
¥ Ground the recorder.
¥ Ground the measuring object.
¥ Insulate TC from the measuring object.
¥ Use shielded cables for input signals.

¥ Check the wiring connection being
correct. (See subsection 2.2.3.)
¥ Attach the terminal block correctly.
¥ Tighten the terminal screws again.
¥ Insulate the RTD from grounding wire.
¥ Replace the TC if its wire is broken.

Is the recorder parallel

connected with other

instruments?

NO

YES

¥ Halt the burnout function of other
instruments.
¥ Make one-point grounding together
with other instruments.
¥ Cancel the parallel connection (for
example, use a double element TC).

Contact OMEGA Engineering.

NO

YES

Is pen position/dot-printing

position correctly adjusted?

Readjust the pen position/dot-printing position. See subsection 4.8.6, 4.8.7.