Appendix C: Specifications

K1297-G20 Monitor User Manual

C-7

Table C–7: Certifications and compliances (Cont.)

Category

Standards or description

U.S. Nationally
Recognized Testing
Laboratory Listing

Only for devices with UL symbol on the back of the housing:

UL 3111-1: Standard for electrical measuring and test
equipment

Canadian
Certification

Only for devices with UL symbol on the back of the housing:

CAN/CSA C22.2 No. 1010.1: Safety requirements for electrical
equipment for measurement, control and laboratory use

Installation (Over-
voltage) Category

Overvoltage Category II (as defined in IEC 61010-1, Annex J)

Local-level mains (wall sockets). Equipment at this level
includes appliances, portable tools, and similar products.
Equipment is usually cord-connected.

Pollution Degree

Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for
indoor use only.

Normally only dry, nonconductive pollution occurs. Occasionally
a temporary conductivity that is caused by condensation must
be expected. This location is a typical office/home environment.
Temporary condensation occurs only when the product is out of
service.

Equipment Type

Test and measuring

Safety Class

Class 1 (as defined in IEC 61010-1, Annex H) – grounded
product

Laser Compliance

All lasers used in this product are in compliance with the
applicable requirements of the following, and are Class I
according to each:
EN60825–1
IEC60825–1
U.S. CDRH Regulations