Section 3—technical description – Activeforever Invacare SOLO2 Portable Oxygen Concentrator User Manual

Page 12

Advertising
background image

SECTION 3—TECHNICAL DESCRIPTION

Invacare SOLO

2

12

Part No 1156872

SECTION 3—TECHNICAL
DESCRIPTION

The Invacare transportable oxygen concentrator is to be used by patients 

with respiratory disorders who require supplemental oxygen. The device 

is not intended to sustain or support life. 
The oxygen concentration level of the output gas ranges from 87% to 

95.6%. The oxygen is delivered to the patient through the use of a nasal 

cannula. The transportable oxygen concentrator provides two modes of 

operation; pulse flow and continuous flow. In pulse flow, when the 

demand for oxygen is detected, the oxygen is delivered with pulse flow 

settings of 1 through 5. In continuous flow, oxygen is continuously flowing 

with settings from 0.5 LPM to 3.0 LPM.
The Invacare transportable oxygen concentrator uses a molecular sieve 

and pressure swing adsorption methodology to produce the oxygen gas 

output. Ambient air enters the device, is filtered and then compressed. 

This compressed air is then directed toward one of two nitrogen adsorbing 

sieve beds. Concentrated oxygen exits the opposite end of the active sieve 

bed and is directed into an oxygen reservoir where it is delivered to the 

patient in specific volumes during the inhalation portion of a detected 

breath.
The Invacare transportable oxygen concentrator is capable of operation by 

the patient in a home environment, in an institutional environment or in a 

vehicle or other mobile environment. Device standard power options 

include an AC to DC switching power supply operating from AC power 

outlet (120 VAC/ 60 Hertz or 230VAC/50 Hertz nominal), a DC supply 

operating from accessory outlets typically found in a mobile vehicle type 

environment (12 VDC nominal) and a rechargeable battery module.