SECTION 6—OPERATING INSTRUCTIONS

Invacare SOLO

2

26

Part No 1156872

Breath Detect Indication
(Pulse Mode Only) The GREEN indicator light will blink off each time an 

inhalation is detected and the transportable oxygen concentrator outputs a 

measured pulse of oxygen.

Audible Alarm
The transportable oxygen concentrator is equipped with an audible alarm. 

The audible alarm will activated in conjunction with Indicators and LCD 

Display to annunciate Alarms and Warnings. 
The transportable oxygen concentrator will also issue a steady beep when 

all power sources are removed. The beeping will stop after 10‐20 minutes or 

once a power source is connected.
The transportable oxygen concentrator will briefly turn on all three lights 

and the LCD display and check for stuck buttons upon power up. It will 

then output a brief audible beep to indicate the unit is operational.

Control/Adjustment Buttons

There are four buttons located surrounding the LCD display. They are used 

to turn the unit on/off, to proceed to the next screen, to select and highlight 

adjustable settings and to adjust those settings.

On/Off Button
To turn the transportable oxygen concentrator on or off, press and hold the 

On/Off button for a period of 2‐3 seconds, not to exceed 5 seconds. Releasing 

the button too soon will not allow the unit to do either. If the button is held 

too long when turning on the unit, the Stuck Button alarm could be 

activated. 

Mode/Return/Highlight Button
Pressing button for 1‐2 seconds in the Standby screen (FIGURE 6.5 on 

page 27) will switch to Normal Operating screen (FIGURE 6.6 on page 27).
Pressing this button for 1 second in the Normal Operating screen will toggle 

between the continuous and pulse modes. 
Pressing this button for 2‐3 seconds in the Normal Operating screen will 

display the LCD Adjustment screen. In the LCD Adjustment screen 

pressing this button for 1 second will switch columns on the screen. To save 

changes and exit the LCD adjustment screen the highlight bar needs to be 

positioned in the left column, then press this button for 2‐3 seconds to return 

to the Normal Operating screen. If the screen switches without pressing and 

holding the button for 2‐3 seconds the changes made in the LCD 

Adjustment screen will not be saved.

Increase and Decrease Buttons
When the highlight bar is on the setting name, these buttons are used to 

select the next setting in the list. When the highlight bar is on the setting 

value, these buttons will change the value to the next value available.