SECTION 8—TROUBLESHOOTING GUIDE

Part No 1156872

41

Invacare SOLO

2

LCD DISPLAY:

INDICATORS:

Continuous audible beep every half

second
RED indicator On.
GREEN indicator is Off.

DESCRIPTION:

SOLUTIONS:

A stuck button is being detected upon

power up.

1. With the unit turned Off, press each

control button looking for a stuck button.
2. Turn On the unit while ensuring that you

are not continuing to hold the Power

On/Off button once the indicator lights and

LCD begin to illuminate.
3. ONLY press the On/Off button during

startup.

ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart. Call your

provider and report the problem if alarm continues.

LCD DISPLAY:

INDICATORS:

Triple audible beep every 10 sec
YELLOW indicator Flashing.

DESCRIPTION:

SOLUTIONS:

Remaining battery capacity has fallen to

15%. Battery requires charging.
Battery gauge is empty.
Loss of power is imminent.

1. Connect transportable oxygen

concentrator to either AC or DC power,

or use an alternate charged battery

module.
2. If no other power source is available,

Change to another source of oxygen if

alternate power is not available.

ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or