SECTION 7—CLEANING, CARE, AND MAINTENANCE

Invacare SOLO

2

32

Part No 1156872

SECTION 7—CLEANING, CARE, AND
MAINTENANCE

ƽ WARNING

Turn Off the transportable oxygen concentrator and unplug the power 

cord before cleaning. 
DO NOT allow any cleaning agent to drip inside the air inlet and outlet 

openings, or the battery pack. 
DO NOT spray or apply any cleaning agent directly to the cabinet.

CAUTION

DO NOT clean the cabinet, carry bag, or filter with alcohol and alcohol 

based products (isopropyl alcohol), concentrated chlorine‐based products 
(ethylene chloride), and oil‐based products (Pine‐Sol

®

, Lestoil

®

) or any 

other harsh chemical agents. Only use mild liquid dish detergent (such as 
Dawn

).

Cabinet

Periodically clean the transportable oxygen concentrator cabinet when 

dirty as follows:
1.

Use a damp cloth, or sponge, with a mild detergent such as Dawn dish 
washing soap to gently clean the exterior case.

2.

Allow the unit to air dry, or use a dry towel, before operating the unit.

Air Intake Filter

CAUTION

DO NOT operate the transportable oxygen concentrator without the air 

intake filter installed. 

NOTE: For this procedure, refer to FIGURE 7.1.

1.

Clean the air intake filter at least once a week depending on 
environmental conditions.

2.

Squeeze the thumb tabs on the inlet filter grill and remove from the 
unit. 

3.

Lift out the filter.

4.

Use a vacuum cleaner or wash with a mild liquid dish detergent (such 
as Dawn™) and water. Rinse thoroughly. 