SECTION 8—TROUBLESHOOTING GUIDE

Part No 1156872

43

Invacare SOLO

2

LCD DISPLAY:

INDICATORS:

Continuous audible beep every half

second.
RED indicator is On.
GREEN indicator is Off.

DESCRIPTION:

SOLUTIONS:

Abnormal operating condition has been

detected

1. Turn unit Off, then On again to retry.
2. Change to another source of oxygen if

alarm continues.

ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart.

Call your provider and report the problem if alarm continues.

LCD DISPLAY:

INDICATORS:

Ten audible beeps every 10 sec
RED
indicator Flashing Fast.

DESCRIPTION:

SOLUTIONS:

Battery pack is too hot, or too cold, to

allow battery to continue to operate.

1. Remove battery and only use AC, or

DC, adapter for power. Or, use an

alternate battery pack.
2. Move battery to warmer, or cooler,

surroundings. Allow battery to cool down

to less than 95° F (35° C), or warm up to

50° F (10° C).
3. Turn unit off. Re-insert battery and

retry.
4. Change to another source of oxygen if

no other power source is available.

ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or

revert to normal operating screens