SECTION 8—TROUBLESHOOTING GUIDE

Invacare SOLO

2

40

Part No 1156872

LCD DISPLAY:

INDICATORS:

Single audible beep every 15 sec
YELLOW
indicator On.

DESCRIPTION:

SOLUTIONS:

(Only possible when External Power

is applied)
Battery is too hot, or cold, to allow for the

battery to charge.

1. Move unit to warmer, or cooler,

surroundings. Allow unit to cool down to

less than 95° F (35° C), or warm up to 50°

F (10° C). Use AC or DC power.
2. Remove battery and only use AC or DC

power.
3. Turn unit Off, then On again to retry.

ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or

revert to normal operating screens.

If the warning is issued and acknowledged a second time, the warning will be

deactivated until the unit is turned off.

Call your provider and report the problem if the warning continues with next use.

LCD DISPLAY:

INDICATORS:

Triple audible beep every 10 sec
YELLOW indicator Flashing.

DESCRIPTION:

SOLUTIONS:

External DC Power is lower than optimal.
Excessive current draw could be present.
Warning will continue every 3 minutes as

long as condition exists.

1. Make sure that car (boat or motor

home, etc.) is running.
2. Make sure DC Power Cable is firmly

connected at both ends.
3. Switch to external AC or battery power.

ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or

revert to normal operating screens