SECTION 8—TROUBLESHOOTING GUIDE

Part No 1156872

45

Invacare SOLO

2

LCD DISPLAY:

INDICATORS:

Continuous audible beep every half

second.
RED indicator is On.
GREEN indicator is Off.

DESCRIPTION:

SOLUTIONS:

Abnormal compressor operation has been

detected.

1. Turn unit Off, then On again to retry.
2. Change to another source of oxygen if

alarm continues.

ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart.

Call your provider and report the problem if alarm continues.

LCD DISPLAY:

INDICATORS:

Triple audible beep every 10 sec
YELLOW indicator Flashing.

DESCRIPTION:

SOLUTIONS:

Oxygen sensor feedback is out-of-range.

Sensor failure is likely

1. Turn unit Off, then On again to retry.

ACKNOWLEDGEMENT REQUIRED:
Press Return / Highlight button. Display will either show additional warnings / alarms or

revert to normal operating screens.

If the alarm is issued and acknowledged a second time, the audible alarm will be

deactivated until the unit is turned off.

Call your provider and report the problem.