SECTION 8—TROUBLESHOOTING GUIDE

Part No 1156872

37

Invacare SOLO

2

Warnings and Alarms

NOTE: In all cases, pressing and holding the On/Off button for 2‐3 

seconds will turn off and reset the unit.

LCD DISPLAY:

INDICATORS:

Triple audible beep every 10 sec
YELLOW indicator Flashing.

DESCRIPTION:

SOLUTIONS:

(Only possible with Pulse Mode)
Transportable oxygen concentrator has

not detected a breath for a predetermined

time period.

1. Verify the cannula is connected, not

kinked, properly positioned and you are

breathing through your nose.

ACKNOWLEDGEMENT REQUIRED:

Press Return / Highlight button. Display will either show additional warnings / alarms or
revert to normal operating screens. This warning will be turned off for a minimum of 5
minutes.

LCD DISPLAY:

INDICATORS:

Single audible beep every 15 sec
YELLOW
indicator On.

DESCRIPTION:

SOLUTIONS:

(Only possible with Pulse Mode)
Your breathing rate has begun to exceed

the capacity of the transportable oxygen

concentrator.

1. Immediately reduce your activity level to

get your breath rate to slow down.
2. Temporarily switch to Continuous Flow

Mode.

ACKNOWLEDGEMENT REQUIRED:

A sufficient reduction of your breath rate will automatically turn off the warning.
Alternately, press Return / Highlight button. Display will either show additional

warnings / alarms or revert to normal operating screens. This warning will be turned off

for a minimum of 5 minutes.