SECTION 8—TROUBLESHOOTING GUIDE

Invacare SOLO

2

46

Part No 1156872

LCD DISPLAY:

INDICATORS:

Continuous audible beep every half

second.
RED indicator is On.
GREEN indicator is Off.

DESCRIPTION:

SOLUTIONS:

External DC Power is too low to allow

continued operation.
Excessive current draw present.

1. Make sure that car (boat or motor

home, etc.) is running.
2. Make sure DC Power Cable is firmly

connected at both ends.
3. Switch to external AC or battery power.
4. Change to another source of oxygen if

no other power source is available.

ACKNOWLEDGEMENT REQUIRED:
Press and hold the On/Off button to power down the unit and restart.