SECTION 6—OPERATING INSTRUCTIONS

Part No 1156872

27

Invacare SOLO

2

LCD Displays

NOTE: For this procedure, refer to FIGURE 6.9 on page 30.

There are four different “Display” screens and two “Adjustment” screens 

available through the LCD display that, along with the Warning and Alarm 

screens, provide full control and information to the user. 

Standby Screen
When the external AC power supply is attached and providing power, the 

unit will show the Standby screen.
When the unit is off and either the Battery or the DC Power are present, a 

momentary push of the On/Off button will display this screen briefly.
This screen is periodically displayed during normal operation. 

FIGURE 6.5 Standby Screen

The Standby screen provides both a battery gauge and a percentage of 

remaining charge. It can also display the Fans Operating, External Power 

On and Battery Charging icons when applicable.

Normal Operating Screen
When operating normally, the Mode & Flow display informs the user of the 

current operating mode and the setting for that mode. 
The display shows the current/last mode of operation selected. The mode of 

operation can be either Continuous Flow or Pulse Dose. It will also show the 

current/last output flow level selected.
This screen will alternate with the Standby screen every 20 seconds. In this 

fashion the user will be able to see all the important information at a glance.

FIGURE 6.6 Normal Operating Screen

Advertising