Cd indicators 2.6 scale base setup (cont.), Procedure – Ohaus CD-33 Indicator User Manual

Page 13

Advertising
background image

13

CD Indicators

2.6 Scale Base Setup (Cont.)

Review the specifications of the scale base to be used with the Indicator. Make sure the settings you select in the
indicator are compatible with the scale base. Alternate displays are shown in this procedure to save space.

Procedure

• Press the Setup button, SCALE is displayed.

ZERO
Zero specifies the percentage of full capacity load (2%,
18% or 100%) that may be cleared by pressing >0<.
Limits may be required in certain applications.

NOTE: 2% Zero capacity is used for hopper scales or
other large scales where accidental zero would lose the
current weight. Bench and counter scales normally use
100% Zero capacity. Legal for trade applications use
2%.

• Press Enter button, ZERO is displayed.

• Press Enter button, ZERO 2% (default) is displayed.

• Press

and select either 2%, 18% or 100%.

• Press Enter button, SAVED is momentarily displayed

then CALUNIT is displayed.

• Press Enter button, CALUNIT kg is displayed.

• Press

and select either kg or lb, then press

Enter button, SAVED is momentarily displayed then
GRAD.

• Press Enter button, GRAD .01 is displayed.

• Press

and select desired setting from .0005

to 100 for scale graduation.

• Press Enter button, SAVED is momentarily displayed

and FULL SC is displayed.

• Press Enter button, FULL SC with a 7 digit number is

displayed, with the first digit flashing. Illustration
indicates a 40 kg scale setting.

• Press

and or

and enter the

desired scale base capacity when using Model CD-31.
The numerical keypad can be used for Model CD-33.
Remember that this number represents either pounds
or kilograms, whatever was previously set for cal unit.

• Press Enter button, SAVED is momentarily displayed

followed by LOCK.