4.1.4 Calibration Test

Calibration test feature allows a check of a known calibration mass against the last stored calibration information in the
indicator. Sample display illustrates a 40 kg scale base using a 10kg mass.

Procedure

• Press the Setup button, SCALE is displayed.

• Press

button UNTIL CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, SPAN is displayed.

• Press

button to select calibration test,CAL

TYPE CALtESt is displayed.

• Press Enter button, TEST WT is displayed and the

last calibration mass value which was entered with the
first digit flashing. (Sample illustrates 10kg).

• Press

and or

and enter the

desired mass value when using Model CD-31. The
numerical keypad can be used for Model CD-33. This
number must be at least 10% of the full span value.

• Press Enter button, CLR PAN is displayed.

• Press Enter button,WORKING is momentarily dis-

played, then PUT WT 10 kg is displayed.

• Place specified calibration mass on pan.

• Press Enter button, WORKING is displayed. After a

few seconds, WEIGH DIFF is displayed.The display
now indicates the actual difference in weight between
the previous weight value which was stored in the
Indicator. After approximately 8 seconds, the display
returns to the WEIGH mode and indicates the weight
on the pan.

• Remove calibration mass from the pan.