Cd indicators – Ohaus CD-33 Indicator User Manual

Page 41

Advertising
background image

41

CD Indicators

Command

Command

Command

Command

Command

Character

Character

Character

Character

Character

Description

Description

Description

Description

Description

?

?

?

?

?

Print current mode: g, kg, custm, %, #

nnnA

nnnA

nnnA

nnnA

nnnA

Set Auto Print feature to "nnnn".
nnnn=0 Turns feature OFF
nnnn=S Output on stability
nnnn=C Output is continuous
nnnn=1-3600 Sets Auto Print interval

C

C

C

C

C

Begin span calibration.

F

F

F

F

F

Print current function.

xl

xl

xl

xl

xl

Set Averaging Filter to "x", where x = 0 to 3.
If LFT, level 0 to 1.
0=minimum level
1=
2=
3= maximum level

L

L

L

L

L

Begin linearity calibration.

xM

xM

xM

xM

xM

Places terminal in mode "x", where x = 1 to 4.
If unit is not already enabled, command returns ES.
1=grams
2=kilograms
3= pounds
4= custom units

P

P

P

P

P

Print display data.

xSL

xSL

xSL

xSL

xSL

Set stable data only printing TO "X", where x = 0 to 3.
If LFT, x=0 or 1.
0=.5d
1= 1d
2= 2d
3=5d

T

T

T

T

T

Same effect as pressing Tare Key.

N

N

N

N

N

Shows copyright notice.

V

V

V

V

V

Print software version number.

EscV

EscV

EscV

EscV

EscV

Print scale ID

xZ

xZ

xZ

xZ

xZ

Set Auto Zero to "x", where x=0 to 3.
If LFT, Auto Zero level from 0 to 1.
0=off
1=0.5d
2=1d
3=3d

x%

x%

x%

x%

x%

Set % reference weight. Uses real number "x" as current Percent Reference Weight.
Weight must be entered in Calibration Units.

x#

x#

x#

x#

x#

Set PC reference weight. Uses real number "x" as current Parts reference Weight. Weight
must be entered in Calibration Units.

EscR

EscR

EscR

EscR

EscR

Resets setup and print menus to factory defaults.
CAUTION: This will reset RS232 configuration.

#

#

#

#

#

Print current Parts Count Reference Weight in the most recently used Units.

%

%

%

%

%

Print current Percent Reference Weight in the most recently used Units.

xF

xF

xF

xF

xF

Set current function to "x", where x=0 to 4.
If mode is not already enabled, command returns ES.
0=None (normal weigh)

RS232 COMMAND TABLE

4.7 Printing Data (Cont.)
4.7.2 Output Formats (Cont.)

Advertising