2 user calibration, Cd indicators 4.1.2 user calibration, Procedure – Ohaus CD-33 Indicator User Manual

Page 30

Advertising
background image

30

CD Indicators

4.1.2 User Calibration

User calibration is used when it is desired to calibrate the Indicator using a mass of known value from 10% to 100% of
the capacity of the scale base. To use this calibration feature, proceed as follows:

(FLASHING)

• Press the Setup button, SCALE is displayed.

• Press

button UNTIL CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, SPAN is displayed.

• Press

button to select user calibration. CAL

TYPE USEr is displayed.

• Press Enter button; the display indicates the last

calibration mass value which was entered with the
first digit flashing. (Sample illustrates 10kg).

• Press

and or

and enter the

desired mass value when using Model CD-31. The
numerical keypad can be used for Model CD-33. This
number must be at least 10% of the full span value.

• Press Enter button, CLR PAN is displayed.

• Press Enter button,WORKING is momentarily dis-

played, then PUT WT 10 kg is displayed.

• Place specified calibration mass on pan.

• Press Enter button, WORKING is displayed. After a

few seconds CAL SET is displayed, the display then
returns to WEIGH mode and indicates mass on pan.

• Remove calibration mass from the pan.

• User calibration is completed.