1 span calibration, Cd indicators 4.1.1 span calibration, Procedure – Ohaus CD-33 Indicator User Manual

Page 29

Advertising
background image

29

CD Indicators

4.1.1 Span Calibration

Span calibration requires a mass equal to 100% of the scale base capacity.

Procedure

• Press the Setup button, SCALE is displayed.

• Press

button until CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, CAL TYPE SPAN is displayed.

• Press Enter button, CLR PAN is displayed. Clear

the pan.

• Press Enter button, WORKING is momentarily dis-

played, then PUT WT is displayed (capacity of base)
on the pan.

• Place specified calibration mass on pan. The illustra-

tion sample indicates 40 kilograms.

• Press Enter button, WORKING is displayed. After a

few seconds CAL SET is displayed; the display then
returns to WEIGH mode.

• Span calibration is completed.

• Remove calibration mass from the pan.