7 legal for trade (lft), Cd indicators 3.7 legal for trade (lft), Procedure – Ohaus CD-33 Indicator User Manual

Page 22

Advertising
background image

22

CD Indicators

3.7 Legal for Trade (LFT)

Legal for Trade (LFT) is a software controlled option which can be set to LFT LOCK. When LFT LOCK is set ON,
certain items in the Calibration, Readout, Print, Mode and Units menus are automatically preset and locked. This is
done to permit the Indicator to operate in a legal for trade application and works in conjunction with a Lock Switch
(jumper). Default setting is UNLOCKED, which limits display resolution to 1:10,000. See default table.

LFT and Lock Switch Menu lock Available Settings

Scale Menu
Zero

Locked

2%

Calibration Unit

Locked

lb or kg

Grad

Locked

1:10,000 max.

Full Scale

Locked

ON

Lock

Locked

ON

Calibration Menu
Span, Linearity, User

Locked

CalTest

Unlocked

Readout Menu
Stability

Unlocked .5d (limited to .5d and 1d)

Auto zero

Unlocked .5d (limited to OFF and .5d)

Filter Level

Unlocked -1-

GMP Data Menu

Unlocked

GMP Selections

Unlocked

Print Options

Unlocked

RS232 Menu

Unlocked

LFT Menu

Lockswitch Locked

Mode menu

Locked

Weigh

Units Menu

Locked

Cal unit*

Global Menu

Locked

Custom Menu

Unlocked

If Print Numeric Data is turned ON, then Print Stable Data Only is
locked ON.

* Any unit which was enabled may be selected as a weigh-
ing unit. Default is the calibration unit. Custom unit is
disabled in LFT and can not be selected as a weighing unit.

DEFAULT TABLE

Procedure

When LFTLOCK is set to ON, None of the settings in the
menus which are listed as Locked in the default table can
be altered.
The settings in the menus which are listed as Unlocked in
the default table may be altered if: A) The Lockswitch
jumper is on (Lockswitch disabled) or B) The Lockswitch
jumper is off (Lockswitch enabled) and the Menulock is off.

When LFTLOCK is set OFF:
The settings of all menus may be altered if: A) The
Lockswitch jumper is on (Lockswitch disabled); or B) The
Lockswitch jumper is off (Lockswitch enabled) and the
Menulock is off.

• Press the Setup button, SCALE is displayed.

• Press

button repeately until LFT is displayed.

• Press Enter button, LFT LOCK is displayed.

• Press Enter button.

• Press

or

button and select either ON or OFF.

• Press Enter button, SAVED is momentarily displayed,

then EXIT is displayed.

• Press Enter button, Indicator returns to weigh mode.

NOTE: For legal for trade applications, the indicator must
be physically sealed. Refer to section on LFT Sealing.

When the Indicator is first turned ON and LFT has been
previously set ON, the following display will appear if LFT
is set in the menu and the Lock Switch is set ON.

When the Indicator is first turned ON and LFT has been
previously set ON, the following display will appear if LFT
is set in the menu and Calibration menu is locked, and
the Lock Switch is enabled (jumper off).

CAL SAFE message appears when Lock Switch is
enabled (jumper off) and Calibration menu lock is ON.