3 linearity calibration, Cd indicators procedure – Ohaus CD-33 Indicator User Manual

Page 31

Advertising
background image

31

CD Indicators

Procedure

• Press the Setup button, SCALE is displayed.

• Press

button UNTIL CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, SPAN is displayed.

• Press

button to select linear calibration. CAL

TYPE Lin is displayed.

• Press Enter button, CLR PAN is displayed.

• Press Enter button, WORKING is momentarily dis-

played, then PUT WT 20kg is displayed.

• Place specified calibration mass on pan.

• Press Enter button, WORKING is displayed. After a

few seconds, display changes to PUT WT 40kg.
The displayed weight is the full capacity of the
Indicator.

• Place specified calibration mass on pan.

• Press Enter button, WORKING, is displayed. After a

few seconds, CAL SET is displayed; the display then
returns to WEIGH mode and indicates mass on pan.

• Remove calibration mass from the pan.

• Linearity calibration is completed.

4.1.3 Linearity Calibration

Linearity calibration utilizes three calibration points; one at zero, center span and full span. This method minimizes
deviation between actual and displayed weights within the Indicator's weighing range. Three weight values are used;
zero, a weight value at or near midpoint of the Indicator's weighing range and a weight value at specified capacity.
Sample display illustrates a 40 kg scale base.