5 calibration gmp printout, Cd indicators 4.1.5 calibration gmp printout, Span calibration printout – Ohaus CD-33 Indicator User Manual

Page 33: Linearity calibration printout, Calibration test printout

Advertising
background image

33

CD Indicators

4.1.5 Calibration GMP Printout

If any option in the GMP Set Menu is turned On, GMP automatically prints data after calibration is completed.

- - - - - SPAN CAL - - - - - -
12/01/99

1:10:00 PM

Scale Id 1234
Cal:

40kg

Old:

39.99kg

Dif:

0.01kg

Wt.
Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

- - - - - LIN CAL - - - - - -
12/01/99

1:20:00 PM

Scale Id 1234
Cal:

40kg

Old:

39.99kg

Dif:

0.01kg

Wt.
Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

- - - - - CAL TEST - - - - - -
12/01/99

1:30:00 PM

Scale Id 1234
Cal:

10kg

Act:

10kg

Dif:

0.00kg

Wt.
Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

Span Calibration Printout

When performing Span calibration with all GMP options
turned on, a printout is automatically made after the
calibration is completed.

Linearity Calibration Printout

When performing a Linearity calibration with all GMP
options turned on, a printout is automatically made after
the calibration is completed.

Calibration Test Printout

When performing a Calibration Test with all GMP options
turned on, a printout is automatically made after the
calibration is completed.

Advertising