4 setup menu, Cw-11 indicator – Ohaus CW-11 Indicator User Manual

Page 14

Advertising
background image

11

CW-11 Indicator

2.8 Initial Setup (Cont.)

2.8.4 Setup Menu

The CW-11 Indicator Setup Menu must be entered the first time the Indicator is used to set the scale base
parameters to match the Indicator. Do not attempt to calibrate the Indicator before setting up the Setup
Menu. All other menus should be entered and set up the first time the Indicator is used.

Procedure

With the Indicator ON, press and hold the G/N/T/
MENU button until MENU is displayed. When you
release G/N/T/MENU button, CAL is displayed when
the CAL jumper on the PC board is in place. When
the CAL jumper is removed, the Indicator will not
permitcalibration. This jumper should be in place
initially.

Press PRINT/UNITS button, SETuP is displayed.

Press G/N/T/MENU button, LFTOFF is displayed.
legal for trade selections are:

'LFT ON' - LFT is ON
'LFTOFF' - LFT is OFF
'LFTCAn' - LFT is set for Canada

Press PRINT/UNITS button and select either ON ,
OFF or Canada.

Press G/N/T/MENU button, 0 2 is displayed. This
is the Zero 2%, 18% or 100% setting. 2% - zero
operation range is - 2% to + 2%. 18% - zero
operating range is -2% to +18%, 100% - zero
operation range is -2% to +100%.
NOTE: If LFT is ON, only 2% and 18% are available.

Press PRINT/UNITS button, and select either 2%,
18% or 100%.

Press G/N/T/MENU button, CAL Un kg is displayed.
This is the calibration unit setting. Selections are:

'lb' - calibration unit is lb

'kg' - calibration unit is kg.

Press PRINT/UNITS button, and select either kg or lb.