Cw-11 indicator – Ohaus CW-11 Indicator User Manual

Page 15

Advertising
background image

12

CW-11 Indicator

Press G/N/T/MENU, F xx is displayed. This is full

scale capacity selections. xx= value last set.

Selections are:

5, 10, 15, 20, 25, 30, 40, 50, 60, 75, 100, 120,

150, 200, 250, 300, 400, 500, 600, 750, 1000,

1200, 1500, 2000, 2500, 3000, 5000, 7500,

10000, 20000 (lb or kg).

Press PRINT/UNITS button until desired capacity

value is reached.

Press G/N/T/MENUbutton, Gd0.01 is displayed.

This is the graduation size. Refer to paragraph

2.8.3 Load Cell Capacity Information table. For

available selections, press PRINT/UNITS button until

desired graduation value is reached.

Press G/N/T/MENU button, CP 30 kg is displayed.

This is the full scale calibration point setting. The

range is from 20% to 100% Full scale capacity.

Press PRINT/UNITS button until desired calibration

value is reached.

Press G/N/T/MENU button, AS 3 is displayed. This

is the duration of animal sample time which can be

set from 1 to 10 seconds.

Press PRINT/UNITS button until desired sample time

value is reached. See Note below.

Press G/N/T/MENU button, Ad 4 is displayed. This

is the duration of animal weighing display time

which can be set from 1 to 10 seconds. The default

is 10 seconds. See Note below.

Press PRINT/UNITS button until desired sample time

value is reached. See Note below.

Press G/N/T/MENU button to end this block, SAVE is

displayed.

Press G/N/T/MENU button, rEAD is displayed which

is the next menu or press PRINT/UNITS button to

return to Setup menu.

The Indicator is now matched up with the scale

base and the Indicator parameters may now be set

and calibrated.

k g

2.8 Initial Setup (Cont.)

2.8.4 Setup Menu (Cont.)

k g

k g

NOTE: Not available if LFT is ON. Only available if ALT
is set to ANI.