11 animal weighing, Cw-11 indicator 4.operation (cont.) – Ohaus CW-11 Indicator User Manual

Page 32

Advertising
background image

29

CW-11 Indicator

4.OPERATION (Cont.)

4.11 Animal Weighing

Animal weighing is enabled only when 'Ani' in the
Read menu under Alternate Mode is turned ON and
animal sample time and animal display time have
been set in the Setup menu. Refer to paragraphs
2.8.4. and 2.8.5. Those menus allows setting animal
sample time up to 10 seconds and to display the
averaged animal weight for up to 10 seconds. Longer
sampling times are used for highly active subjects.

With the Indicator in a weighing mode, place a
holding container if used on the base. Press the
TARE button to tare the container weight.

Place the subject in the container and momentarily
press the PRINT/UNITS button. The display counts
down the number of seconds you have set, aver-
ages the subjects weight and then displays this for
the number of seconds you have selected. The
average weight shown is stable. The value is also
sent to the RS232.

At the end of the time out period, the actual weight is
shown as the Indicator reverts to a normal weighing
mode. If the subject is moving a lot, the displayed
weigh will fluctuate.

Press the TARE button to remove the container
weight from memory unless you intent to weigh
additional subjects using the same container. The
Indicator is now in a normal weighing mode.

k g

*

Gross

Brutto

Center

of Zero

Sample Container Weight

k g

*

Gross

Brutto

Tared Container weight

k g

*

k g

*

Displayed Average Animal Weight

Net

Dynamic

Advertising