CW-11 Indicator

2.8 Initial Setup (Cont.)

2.8.5 Readout Menu (Cont.)

Procedure

AVERAGING LEVEL (Cont.)

Press G/N/T/MENU button, AL LO is displayed. This is
averaging level settings. Selections are:

'Lo' - Averaging level is low
'Hi' - Averaging level is high.

Press PRINT/UNITS button,and select LO or HI.

STABILITY

The stability range specifies the weighing results and
must be within a preset tolerance limit for a certain
time to turn the stability indicator ON. When a dis-
played weight changes beyond the allowable range,
the stability indicator turns OFF, indicating an unstable
condition. Factory default setting is shown in bold
type.

.5d Smallest range: stability indicator is

ON only when displayed weight is

within .5 divisions.

1d Normal setting. - Fixed for LFT

3d More stable course

Press G/N/T/MENU button, StAb1 is displayed.The
stability range specifies the weighing results and
mustbe within a preset tolerance limit for a certain
time to turn the stability indicator ON. When a
displayed weight changes beyond the allowable
range, the stability indicator turns OFF, indicating an
unstable condition. 0.5 d smallest range, stability
indicator is ON only when displayed weight is stable
within 0.5 divisions. 1 d-stable within 1 division. 3
d-largest range, stability indicator is ON even
though displayed weight changes 3 divisions.
Factory default setting is 1.

Press PRINT/UNITS button,and select 0.5, 1, or 3.
Normal 1d stability is default/recommended.