13

CW-11 Indicator

2.8 Initial Setup (Cont.)

2.8.5 Readout Menu

The Readout menu is used to adapt the Indicator to environmental conditions, set measuring units on/off, alter-
nate modes - animal weighing & parts counting, auto zero tracking, timer on/off, retain zero data and backlight-
ing. Review all of the settings available before proceeding.

Procedure

To select any of the items in the Readout menu,
proceed as follows: NOTE: If you have entered from
the preceeding menu, disregard the first step.

With the Indicator ON, press and hold the G/N/T/
MENU button until MENU is displayed. When you
release the G/N/T/MENU button, CAL is displayed,
then press PRINT/UNITS button, until rEAd is dis-
played.

Press G/N/T/MENU button, rESETn is displayed.
This allows resetting the readout menu to factory
defaults. rESETn = no and does not reset settings.
rESETy= yes and will reset the entire readout menu
as follows: AL Lo, StAb 1, Un Off g, Un On kg, Un Off,
lb, Un Off oz, Alternate Mode Off, AZt 0.5, Aot Off, rZd
Off and Backlight auto.

Press PRINT/UNITS button,and select N or Y.

AVERAGING LEVEL

Averaging level compensates for vibration or exces-
sive air currents on the scale base. During operation,
the indicator continually takes weight readings from
the load cell. Successive readings are then digitally
processed to achieve a stabilized display. Using this
feature, you specify how much processing you need.

HI and LO settings are available.
HI setting:
More processing, greater stability and
slower stabilization time.

LO setting:
Less processing, less stability and faster
stabilization time.