Cw-11 indicator – Ohaus CW-11 Indicator User Manual

Page 24

Advertising
background image

21

CW-11 Indicator

Procedure (Cont.)

With the Indicator ON, press and hold the G/N/T/

MENU button until MENU is displayed. When you

release the G/N/T/MENU button, CAL is displayed,

then press PRINT/UNITS button, until LOCSW is

displayed.

Press G/N/T/MENU button, LSTOFF is displayed.

This permits locking the Setup menu. OFF is un-

locked, ON is read only (locked). This menu is

hidden if the CAL jumper is off.

Press PRINT/UNITS button, and select ON or OFF.

Press G/N/T/MENU button, LrdOFF displayed. This

permits locking the Readout menu. OFF is un-

locked, ON is read only (locked).

Press PRINT/UNITS button, and select ON or OFF.

Press G/N/T/MENU button, LPtOFF is displayed.

This permits locking the Print menu. OFF is un-

locked, ON is read only (locked).

Press PRINT/UNITS button, and select ON or OFF.

Press G/N/T/MENU button, LCLOFF is displayed.

This permits locking the Calibration menu. OFF is

unlocked, ON is read only (locked). This menu is

hidden if the CAL jumper is off.

Press PRINT/UNITS button, and select ON or OFF.

Press G/N/T/MENU button to end this block, SAVE

is displayed.

Press G/N/T/MENU , Quit is displayed..

Press PRINT/UNITS button to go to CAL or press

G/N/T/MENU button, Indicator returns to a weighing

mode.

NOTE: At this point, the Indicator must be calibrated

and the jumper removed from the CAL connector in

order to lock out the menus. The top cover of the

Indicator should be free to gain access to the CAL

jumper.

Center

of Zero

Gross

Brutto

2.8 Initial Scale Setup (Cont.)

2.8.7 Lockout Switch Menu (Cont.)

k g