Cw-11 indicator – Ohaus CW-11 Indicator User Manual
Page 28
25
CW-11 Indicator
Replace the battery cover and four battery cover
screws. One of these screws is cross drilled and
can accept a wire seal.
NOTICE: The CW-11 has been tested and found to
comply with Class lll requirements of NIST Handbook
44.
After the Indicator has been tested and found to
comply with local applicable regulations by a local
weights and measures official, it may be sealed as
follows:
LEAD AND WIRE SEAL
See illustration at left. Place wire seal through the
holes in the screw and ribs as shown and com-
press lead seal in place.
PAPER SEAL
If an audit trail or paper seal will be used, install the
6-32 pan head screw to the battery cover and place
seal over the screw area.
3.1 Legal for Trade (LFT) Operation and LFT Sealing (Cont.)
SEALING SCREW
BATTERY COVER
Sealing the Indicator
CLOSING THE INDICATOR
After the Indicator has been calibrated and sealed,
replace the rear cover and snap into place. Connect
power to the Indicator and verify operation.
SCREW
BATTERY HOLDER
PAPER SEAL