Cw-11 indicator – Ohaus CW-11 Indicator User Manual

Page 19

Advertising
background image

16

CW-11 Indicator

AUTO ZERO

Press G/N/T/MENU button, AZt 0.5 is displayed.
This is the Auto Zero Threshold setting. Auto Zero
minimizes the effects of temperature changes and
small disturbances on the zero reading. The
Indicator maintains the zero display until the thresh-
old is exceeded. Settings are shown as follows:

OFF
0.5d Sets threshold to 0.5 divisions. - Fixed in LFT

1d Sets threshold to 1 division.
3d Sets threshold to 3 divisions.

Factory default setting is 0.5d.

Press PRINT/UNITS button,and select either 0.5, 1 or
3.

AUTO POWER OFF

Press G/N/T/MENU button, AOtOFF is displayed. This
is the Auto Off Timer. When set ON, the Indicator will
shut off automatically after 5 minutes has elapsed
based on the condition that no button is pressed
and the scale base is stable during that period.

Press G/N/T/MENU button,and select ON or OFF.
OFF is the default setting.

RETAIN ZERO DATA

Press G/N/T/MENU button, Un rZdOFFis displayed.
This is Retain Zero Data which can be turned on or
off. When set On, the Indicator stores the current
zero point and restores it on the power-up.

Press PRINT/UNITS button, and select ON or OFF.
OFF is the default setting.

2.8 Initial Setup (Cont.)

2.8.5 Readout Menu (Cont.)

Advertising