Types and versions – VEGA VEGASON 51K…53K User Manual

Page 11

Advertising
background image

VEGASON 51K … 53K

11

4 … 20 mA

2

2

2

2

4

Measuring system with VEGASON 51K … 53K on active DCS

• Two-wire technology, supply on active DCS.
• Output signal 4 … 20 mA (passive).
• Measured value indication integrated in the sensor.
• Optional external indicating instrument (can be mounted up to 25 m away from the sensor

in Ex-area).

• Adjustment with PC, HART

®

-handheld or adjustment module (can be plugged in the sensor

or in the external indicating instrument).

HART

®

-handheld

VEGA-
CONNECT 2

Types and versions

1)

If the resistors of the processing systems
connected to the 4 … 20 mA-signal output are less
than 200

, a resistor of 250

to 350

must be

connected to the connection line during adjust-
ment.
The digital adjustment signal would be extremely
damped or short-circuited via too small input
resistors of the connected processing system
which means that the digital communication with
the PC could no longer be ensured.

2)

4 … 20 mA passive means that the sensor takes,
depending on the level a current of 4 … 20 mA.
The sensor therefore reacts electrically like a
resistor (consumer) on the DCS.

1 )

VEGADIS 50

passive

2)

DCS

Advertising