Diagnose

7 Diagnosis

7.1 Simulation

To simulate a certain filling, you can call up
the function "Simulation" on the adjustment
module MINICOM, in the software program
VVO or in the HART

®

handheld.

With this function, you simulate a and hence a
real sensor current vessel filling level. Please
note that connected instruments, such as
e.g. a PLC, react according to their adjust-
ments and will probably activate alarms or
system functions.

7.2 Error codes

Error codes

Corrective measure

E013

No valid measured value

Message is displayed during the warm-up

- Sensor in the warm-up phase

phase

- Loss of the useful echo

If the message remains, a false echo storage
(menu "Sensor optimisation") must be carried out
with the adjustment modul MINICOM or better
with the PC and VVO.
If the meassage is still displayed, you have to
carry out a readjustment.

E017

Adjustment span too small

Carry out a readjustment.
Make sure that the difference between
min. and max. adjustment is at least 10 mm .

E036

Sensor software does not run

Sensor requires a software (service).
Message appears during a software update.

E040

Hardware failure/Electronics defec-

Check all connection cables.

tive

Transducer defective.
Contact our service department.