VEGA VEGASON 51K…53K User Manual

Page 19

Advertising
background image

VEGASON 51K … 53K

19

Technical data

Ambient conditions

Max. vessel pressure (gauge pressure)
- VEGASON 51 and 52

3 bar

- VEGASON 53

1 bar

Ambient temperature
- sensor (electronics)

-20°C … +60°C

- process (transducer)

-40°C … +80°C (StEx: -20°C … +75°C)

- storage and transport temperature

-40°C … +80°C

Protection
- sensor

IP 67

- transducer, process

IP 68

Protection class
- two-wire sensor

II

- four-wire sensor

I

Overvoltage category

III

Self-heating
at 40°C ambient temperature
- to sensor

45°C

- to transducer, process

55°C

Ex-technical data

Classification

ia intrinsically safe (in conjunction with a safety
barrier or separator)

Temperature class (permissible ambient
temperature on the transducer when used
in Ex-areas)
- T6

42°C

- T5

58°C

- T4

60°C

- T3

60°C

Ex-approved in category or zone
- ATEX

Zone 1 (II 2 G)

- IEC, CENELEC, PTB

Zone 1 (II 2 G)

Classification mark

EEx ia IIC T6

Process connections

VEGASON 51

G 1

1

/

2

A, 1

1

/

2

" NPT

VEGASON 52

G 2 A, 2" NPT

VEGASON 53

DN 100 compression flange

Connection lines

Two-wire sensors

supply and voltage via one two-wire line,
load dependent on the supply voltage

Four-wire sensor

supply and signal separately, load of the
4 … 20 mA-signal line max. 500

Cross-section area of conductor

generally 2,5 mm

2

Earth connection

max. 4 mm

2

Cable entry, Pg

2 x M20 x 1,5 (cable diameter 5 … 9 mm)

Advertising