4 adjustment with the hart®-handheld, 4 adjustment with the hart, Handheld – VEGA VEGASON 51K…53K User Manual

Page 46

Advertising
background image

46

VEGASON 51K … 53K

Connection to a VEGA-signal condi-
tioning instrument

When you operate a HART

®-

capable sensor

on a VEGA-signal conditioning instrument,
you have to connect the sensor via a resistor
according to the following table during the
HART

®

-adjustment, to reach together with the

inner resistor of the instruments the value of
250 Ohm required for the HART

®

-instrument.

VEGA-signal conditioning instr.

Rx

VEGAMET 513, 514, 515, 602

50 … 100 Ohm

VEGAMET 614

no additional

VEGADIS 371

resistor
required

VEGAMET 601

200 … 250 Ohm

VEGASEL 643

150 … 200 Ohm

VEGAMET 513 S4, 514 S4
515 S4, VEGALOG EA-card

100 … 150 Ohm

Set-up

With each HART

®

-handheld you can set-up

the VEGASON series 50K ultrasonic sensors
like all other HART

®

-capable sensors. A spe-

cial DDD (Data Device Description) is not
necessary.

Just connect the HART

®

-handheld to the

signal line, after having connected the sensor
to power supply.

Note:

If the resistor of the power supply is less than
250 Ohm, a resistor must be looped into the
signal/connection line during adjustment.

The digital adjustment and communication
signals would be short-circuited via too small
resistors of the supply current source of the
processing system which means that the
sensor communication could not be ensured.

Ri < 250

+

-

250

Ri < 250

+

-

6.4 Adjustment with the HART

®

-handheld