26

VEGASON 51K … 53K

Mounting boss on dished tank ends

Open vessels

Reference
plane

60 mm

1

/

2

vessel radius

Min. meas.
distance

Open vessles

On open vessels the use of the instruments
on a mounting lever is recommended. Mount
the low-weight sensor to a mounting lever.

Flange on dished vessel ends

Reference plane

1

/

2

vessel radium

Mounting and installation

Mounting on short 1

1

/

2

“ or 2“ socket piece

Reference plane

60 mm

You can also mount the sensors with 1

1

/

2

“ or

2“ thread to short socket pieces.

Dished tank ends

On dished tank ends, please do not mount
the instrument in the centre but approx. 1/2
vessel radius from the centre. Dished tank
ends can act as paraboloidal reflectors for
the ultrasonic pulses. If the transducer is
placed in the "focus" of the parabolic tank
end, the transducer receives amplified false
echoes. The transducer should be mounted
outside the "focus". Parabolic amplified ech-
oes are thereby avoided.

60 mm

Reference plane