VEGA VEGASON 51K…53K User Manual

Page 15

Advertising
background image

VEGASON 51K … 53K

15

2

-

+

4 … 20 mA

2

2

2

4

Measuring system with VEGASON 51K … 53K in four-wire technology

• Four-wire technology, supply and output signal via two separate two-wire lines
• Output signal 4 … 20 mA active
• Optional external indicating instrument with analogue and digital indication (can be mounted

up to 25 m away from the sensor)

• Adjustment with PC, HART

®

-handheld or adjustment module MINICOM (can be plugged in

the sensor or in the external indicating instrument VEGADIS 50)

• Max. resistance on the signal output (load) 500

Types and versions

(active)

2)

VEGADIS 50

1 )

1)

If the resistors of the processing systems
connected to the 4 … 20 mA-signal output are less
than 200

, a resistor of 250

to 350

must be

connected to the connection line during adjust-
ment.
The digital adjustment signal would be extremely
damped or short-circuited via too small input
resistors of the connected processing system
which means that the digital communication with
the PC could no longer be ensured.

2)

4 … 20 mA active means that the sensor delivers
a current of 4 … 20 mA (source) depending on the
level. The sensor, therefore, reacts electrically
against a processing system (e.g. indication) like a
current source.

HART

®

-handheld

VEGA-
CONNECT 2

>

200

Advertising